Section 10.12.080 Standards and Specifications for Exhaust Gas Analyzers and Span Gases for Equipment Used to Perform Enhanced Basic Testing.
Prior to the adoption of Ord. 03-2002 on 10/01/2002, Section 10.12.080 read as follows.
A. DC98UTAH2000 exhaust gas analyzer.: specifications
for the DC98 UTAH2000
analyzer are contained in Section 10.12.240 of this Ordinance.
1. Emissions inspections required by this
chapter Ordinance for vehicles subject to enhanced
basic I/M inspections shall be performed using only instruments approved by the Division.
the
UTAH2000 analyzer.
2. Any analyzer used by an enhanced basic I/M program
station shall be registered with and
approved by the Division. and, if required, shall bear a registration sticker issued
by the Division.
Registration stickers are not transferable or assignable. Any new or used exhaust gas analyzer put
in use after enhanced basic I/M program station approval must be approved by the Division
before use. Analyzers used temporarily during times of breakdown or repair of the registered
analyzer do
not require a registration sticker but shall meet all other requirements of this Section including
the approval of the Division before use.
a. The analyzer printers shall be maintained in such
a manner that the printing of the
certificate and inspection report shall be clearly visible on all copies. If any printer
fails to
properly function, then the enhanced basic I/M program station shall discontinue testing until the
required repairs have been performed.
3. Propane Equivalency Factor (P.E.F.). Each
instrument shall be labeled with a valid
propane equivalency factor, shown with an accuracy of at least two decimal places, (i.e., 0.52).
P.E.F. confirmation shall be made on each assembled analyzer by measuring both N-hexane and
propane values on assembly line quality checks. If the analyzer bench is replaced, then a new
P.E.F. label applicable to the replacement bench shall be appropriately attached to the analyzer.
4.3. Running changes.: Any
changes to the design characteristics or component
specifications that may affect the performance of an exhaust gas analyzer to be used as an official
test instrument in the Davis County I/M program shall be approved by the Division. It shall be
the instrument manufacturer s responsibility to confirm that the changes have no detrimental
effect on the performance of the exhaust gas analyzer.
a. It shall be illegal for any person to modify the hardware
or software of an approved
emissions analyzer without written application and formal written approval by the Division.
5. Required Upgrade of the UTAH91 Analyzer. Beginning
January 1, 1998, All UTAH91
analyzers continuing to be used in the DC98 program and all analyzers brought into the program
shall be upgraded with such hardware and software changes as specified by the Division as to
make them capable of real time data transmission, nitrogen oxide emission control equipment
function testing data management, upgraded bench and printing capability and other upgrades
deemed necessary by the Division. Upgraded analyzers will then be called DC98 analyzers.
6.4. Calibration/span gases.
a. General.: The
instrument manufacturer and/or his designated marketing vendors shall, on
request, supply at a reasonable cost, span gases approved by the Division to any ultimate
purchaser of his unit. Each new or used instrument sold by the instrument manufacturer or
marketing vendor shall have approved full span gas containers installed and operational at the
time of delivery. The Division shall establish necessary procedures for approving
span/calibration gases.
b. Span gas blends.: The
span gases supplied to any enhanced basic I/M program station
shall conform to the BAR97 specifications of the Division.
All span/calibration gases shall meet
all Federal requirements for the emissions warranty coverage. Two span/calibration gases shall
be used to perform analyzer calibration. Both a low range and a mid to high range gas approved
by the Division shall be used. Only gas blends supplied by Division approved blenders shall be
used to calibrate official I/M test analyzers.
7.5. Documentation, logistics,
and warranty requirements.
a. An instruction manual shall accompany each exhaust
gas analyzer and shall contain at
least the following information for the analyzer:
i. A complete technical description;
ii. The functional mechanical and electrical
schematics;
iii.ii. The accessories and options
that are included and/or available;
iv.iii. The model number, identification
marking and location;
v.iv. Operating maintenance including
daily, weekly, and monthly, accommodations and
procedures for maintaining sample system integrity including, but not limited to, leaks, hang-up,
calibration and filters. The services to be performed only by the manufacturer shall be clearly
identified;
vi. Field calibration procedures (i.e., Division
inspection procedure with separate gas
supply);
vii. Cal-port gas inlet calibration, zero, and
span instructions;
viii.v. Information concerning
the nearest service facility where equipment can be serviced;
and
ix.vi. The warranty provisions
for the analyzer, including a list of warranty repair facilities
by name, address and telephone number.
x.vii. The analyzer shall be
maintained in accordance with the manufacturers recommended
maintenance schedule and records of this maintenance service shall be maintained for
examination by the Division.
B. The Division shall use and require for use
in the calibration of exhaust gas analyzers,
calibration/span gases and containers meeting the guidelines contained in subsection (A)(6) of
this section.
C.B. Gas span and leak check.
1. A certified emissions repair technician or certified
emissions tester shall perform a gas
calibration of the exhaust gas analyzer, with an approved calibration/span gas, within three (3)
days prior to performing any emissions test, and a leak check within twenty-four (24) hours prior
to performing any emissions test. The gas calibration and leak check must be performed in
accordance with the analyzer specifications as referenced in Section 10.12.240. as prompted
by
the UTAH2000 analyzer.
2. All gas calibration and leak check records
shall be retained and made available to the
Division representatives for review and verification that the appropriate gas span and leak checks
are being performed.
3.2. The analyzer instruction
manual and other Division approved information shall be
reviewed by the certified emissions repair technician or certified emissions tester to ensure that
proper procedures are being used for performing the gas calibration check.
D. At any time where the analyzer requires repairs
in order to be used in accordance with
this chapter, the enhanced basic I/M program station shall notify the Division that the analyzer is
out-of-service and shall again notify the Division after repairs are made and before testing
is
resumed.
E. An NCR type carbonless two-part
eight-inch by eleven (11) inch form shall be used on
the non-dedicated printer for printing the inspection report and other reports required by the
Division unless otherwise determined by the Division. (Ord. 7-1996 § 1 (part))